kSA Emissometer
Ex Situ Wafer Carrier
Characterization
The kSA Emissometer is designed to quickly and easily generate high-resolution reflectance, emissivity and pocket-depth maps of wafer carriers. Wafer carrier emissivity variation can mean temperature nonuniformity, which can lead to reduced device yield and possibly complete growth run failure. With the kSA Emissometer, emissivity changes can be tracked to determine carrier end-of-life without wasting growth runs. The kSA Emissometer also detects residual deposits after baking, and identifies carrier surface defects, scratches, microcracks and pits.
Measures: |
---|
Diffuse Reflectance |
Specular Reflectance |
Emissivity |
Pocket Depth |
Photoluminescence (optional) |
Applications: |
---|
Wafer carrier emissivity and texture uniformity |
Quantitative temperature set-point correction, due to change in wafer carrier emissivity |
Detection of residual material deposits |
Identification of changes in pocket depth and profiles |
Carrier end-of-life determination |
kSA Emissometer Measurement Technology
kSA Emissometer: The kSA Emissometer uses a triple-sensor head for simultaneous height, diffuse reflectance and specular reflectance measurements as the wafer carrier rotates. The head moves on a linear scanning stage radially outward from the center of the carrier, acquiring data at user-specified radial steps to produce a real-time map of the carrier depth, diffuse reflectance, specular reflectance, and emissivity. The system features easy loading and unloading of carriers. An alignment laser allows alignment of carriers to a dimple or other carrier feature for reproducibility from scan to scan. The software interface is designed for operator or engineer use.
Scanning: The kSA Emissometer utilizes a high-resolution linear stage and fast data acquisition electronics to acquire complete carrier maps. Typical maps take measurements every 0.5° in 1mm radial steps, but the angular and radial resolution are user-programmable.
Depth: The depth of the carrier is measured via triangulation. The measured carrier depth is used to correct reflectance measurements for varying optical-head-to-carrier distances. It also provides carrier depth profiles for monitoring incoming wafer carriers and changes in carrier depth profiles with subsequent depositions.
Diffuse Reflectance: The diffuse reflectance of the carrier is measured from light incident at a 45° angle to the carrier and collected at normal incidence. The diffuse reflectance is the main component used to calculate the surface emissivity.
Specular Reflectance: The specular reflectance of the carrier is measured from light incident at a 45° angle to the carrier and collected at the corresponding 45° specularly reflected angle. The specular reflectance is typically small for SiC-coated wafer carriers. However, careful collection of the specular reflection signal provides details of surface imperfections like scratches, pits, and machining marks.
Emissivity: The point-by-point emissivity of the carrier is determined using the measured reflectance values at each point, i.e. by summing the diffuse and specular components of the reflectance. Variation in emissivity across the carrier can lead directly to temperature nonuniformity across the wafers.
UV Photoluminescence (PL): Characterization of MOCVD wafer carriers can be extended with the addition of an UV PL system to the kSA Emissometer. This option uses a LED focused at the wafer carrier and a filtered photodetector focused at the excitation point. The UV PL measurement tracks concurrent with the reflectivity/emissivity/height measurements and maps any features of residues and contaminants to the same positions mapped out with the standard analyses. Ask your k-Space representative to quote this option if you are interested in these capabilities.
kSA Emissometer Functional Specifications
Specifications |
|
Carrier Diameter | Up to 800 mm |
Carrier Compatibility | Spindle centering, works with most commercially-available carriers |
Scanning Stage Resolution | 0.2 mm radial, 0.1° azimuthal, user-programmable |
Scan Time | 10 minutes (for a 0.1° and 1 mm resolution scan of a 465 mm carrier) |
Reflectance / Emissivity |
|
Detection Spot Size | 1.5 mm |
Light Source Wavelength | 940 nm |
Reflectance Resolution | ± 0.005 |
Reflectance Accuracy | ± 0.01 |
Emissivity Resolution | ± 0.01 |
Emissivity Accuracy | ± 0.02 |
Depth |
|
Detection Spot Size | 0.2 mm |
Light Source Wavelength | 660 nm |
Depth Resolution | ± 10 μm |
Depth Accuracy | ± 10 μm |
700mm wafer carrier maps showing Platen Diffuse Reflectance, Platen Emissivity, and Platen Height with a
31-wafer template.
kSA Emissometer Integrated Software
The kSA Emissometer integrated software allows for easy control of the scan and in-depth analysis of the diffuse reflectance, specular reflectance, emissivity, and depth profile data. Data is easily exported to common data file types, as well as to common database formats (XML and JMP).
1 | Partial diffuse reflectance carrier map. The diffuse reflectance is the main contributor to the emissivity calculation. |
2 | Partial specular reflectance carrier map. The specular reflectance is very sensitive to surface defects. |
3 | Partial emissivity carrier map. Note the emissivity variation across the wafer pockets. |
4 | Current carrier emissivity value at the current scan position, shown in real-time during the scan. |
5 | Scanning acquisition parameter dialogue. Enter the Run name, inner and outer radius values and the scan step size. Also select a user-defined analysis template for immediate carrier characterization results. |
kSA Emissometer software screen shot showing scanning dialogues and real-time data acquired from a wafer carrier scan.
kSA Emissometer Analysis Features
Line and Ring Profile Capability:
Users can define line and ring profiles on wafer carrier maps to extract data for further analysis. In the example below, height profiles are displayed along two radial lines from the center to the edge of carrier through a wafer pocket (at 30°) and solely on the carrier web (at 352°). Another application is using ring profiles to plot emissivity data at particular watch radii for comparison with temperature monitoring data.
Web and Pocket Analysis:
User-defined wafer carrier templates may be used to automatically generate a Platen/Carrier Report at the completion of the scan. A 5×6” wafer template is shown in the above carrier map, with each pocket labeled 1-5. This analysis tool determines pocket and web measurement statistics over the entire web and pocket regions and for the individual pocket regions. The user may also define exclusion regions for this analysis as well as specify acceptance criteria for the measured parameter. The data below provides the emissivity statistics for the carrier shown above.
kSA Emissometer System Specifications
System Housing
The kSA Emissometer is a standalone unit as shown below. The lid opens and the carrier rotational stage extends for easy carrier loading and unloading. The lid is closed during data collection to ensure high-quality measurements.
Controller
The kSA Emissometer is supplied with a fully configured, Windows 10, multi-core CPU computer system mounted in the Emissometer enclosure. Supplied 22” LCD Monitor, keyboard and mouse sit outside the enclosure on an adjustable monitor arm. Please refer to the kSA Computer Product Specifications for the latest computer configuration. k-Space does not recommend customers substitute or update their own computer for the controller.
Facility Requirements
Positioning: Four leveling casters are positioned on the bottom of the Emissometer. The unit can be rolled into place and then parked by lowering the feet so the casters are raised off the floor.
System Power: 208 ‐240 VAC, 50/60Hz, single phase, 5 A max, with three wire for box mounting or user specified electrical connector.
System Dimensions: Width=119cm, Depth=102cm, and Height=191cm with the lid closed and Height=241cm with the lid open.
System Weight: 550kg (1212lbs).
Installation and Training
A minimum of 1-2 days of on-site customer installation and training are required.
Warranty
All kSA systems and integrated components are warranted against defective materials and workmanship for a period of ONE YEAR from the date of delivery to the original purchaser.
Distributors
HEADQUARTERS
k-Space Associates, Inc.
Michigan, USA
www.k-space.com
[email protected]
DISTRIBUTION PARTNERS
RTA Instruments Ltd.
Europe
www.rta-instruments.com
[email protected]
El Camino Technologies Pvt Ltd.
India
www.elcamino.in
[email protected]
Giant Force Technology Co., Ltd.
China
www.giantforce.cn
[email protected]
Jung Won Corporation
South Korea
www.jwc.co.kr
[email protected]
R-DEC Co.,Ltd.
Japan
Hong Kong
Taiwan
www.rdec.co.jp
[email protected]